6. COMPARATOR
The results of the comparison are indicated by HI, OK or LO on the display.
The comparison is as follows:
LO < Lower limit value OK Upper limit value < HI
Operating conditions (see the function setting “Cp”):
No comparison (comparator function disabled).
Compares all data.
Compares all stable data.
Compares plus data except those near zero (plus data greater than +4d).
Compares stable plus data except those near zero (stable plus data
greater than +4d).
Compares all data except those near zero (all data greater than +4d or
less than -4d).
Compares stable data except those near zero (stable data greater than
+4d or less than -4d).
d = the smallest display division
e.g.: 4d=four display divisions
The upper limit and lower limit numerical values are common to each of the weighing,
counting and percent mode. An example for the EK-1200i/2000i/3000i is as follows.
Upper limit value “001010”: “101.0g” “1010pcs” “101.0%”
Lower limit value “000990”: “99.0g” “990pcs” “99.0%”
6-1. Setting example
This example will be “Compares plus data except
those near zero”.
Selecting a comparison mode
1. Press and hold the SAMPLE key to display
func .
(If the comparison mode is already set, press the
SAMPLE key to go to “Entering the upper and lower limit values”.)
2. Press the PRINT key, then the balance displays
poff X .
3. Press the SAMPLE key several times to display
Cp X .
4. Press the RE-ZERO key several times to display
Cp 3 .
5. Press the PRINT key to store the settings.
Cp Hi appears after end .
func
poff 0
Cp 0
Cp 3
Press and hold
Press
several times
Press
several times
To store