AD-4212C-300 AD-4212C-600 AD-4212C-3000 AD-4212C-6000 AD-4212C-301 AD-4212C-3100 Production Weighing Unit INSTRUCTION MANUAL 1WMPD4001992B
8Rear of theweighing unitRS-232C cableD-sub 9-pinArrow markface up2.2. Installing the Balance Install the balance as follows: 1. Refer to “3. PRECAU
9MASS OF THE LOCALLY FABRICATED WEIGHING PAN Design the weighing pan so that the mass falls within the ranges shown in the table below: Model Statu
103. PRECAUTIONS To get the optimum performance from the balance and acquire accurate weighing data, note the following: 3.1. Before Use Install
113.2. When Building Into a System The AD-4212C is a precision balance. When it is built into a system, errors such as unstable weight values may o
12 Errors due to a static charge and measures to take When the ambient humidity is less than 45% RH, insulators such as plastic or glass are prone
13 Errors due to air flow and measures to take Where the influence of ambient air flow is great such as: close to an air conditioner, door or pass
143.3. During Use To minimize the affect by electrical noises, earth ground the weighing unit and the display unit (option). Do not drop things
154. OPERATION OF WinCT-AD4212C The software has the following functions: - Set the weighing speed - Set the minimum display value - Calibration
164.1. Changing the Weighing Speed Using WinCT-AD4212C The weighing speed can be selected from the following three rates to minimize the influence
17 Calibration procedure This function calibrates the balance using the calibration weight. (Display example: AD-4212C-300) Operation 1. Plug in t
© 2011 A&D Company, Limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into
184.3. Changing the Baud Rate Using WinCT-AD4212C Operation 1. Click the [Settings] tab. 2. Select the baud rate to be used and click the Set
195. SMART RANGE FUNCTION The AD-4212C-301/3100 are equipped with two ranges. The precision range has a higher resolution. The standard range has a n
206. COMMUNICATION SPECIFICATIONS The AD-4212C series can communicate interactively using RS-232C. The AD-4212C series can continuously transmit th
212. Pin position Pin position of the RS-232C cable (D-Sub 9-pin side, female) Pin No. Signal name Direction Description 1 (Vs) - Internally u
224. Command The balance can be controlled by the following commands. When transmitting, add the terminator <CR><LF> to each command. Co
23 CAL command example (PC, PLC)External devices Time Weighing unit CAL commandDisplays
24 SMP command example R command example when the precision range is used SMP commandSMPCRLFST,+0000.012 gS
257. MAINTENANCE Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair. Use the original pac
268. TROUBLESHOOTING 8.1. Checking the Balance Performance and Environment The balance is a precision instrument. When the operating environment or
27 Sample and container Has the sample absorbed or lost moisture? Has the temperature of the sample, and the container if used, been allowed to
1CONTENTS 1. INTRODUCTION...2
289. SPECIFICATIONS AD-4212C-300 AD-4212C-600 AD-4212C-301 Weighing capacity 320 g 620 g 320 g / 51 g*1 Maximum display 320.084 g 620.084 g
29 AD-4212C-3000 AD-4212C-6000AD-4212C-3100 Weighing capacity 3200 g 6200 g 3200 g / 510 g*1 Maximum display 3200.84 g 6200.84 g 3200.84 g / 51
3010. EXTERNAL DIMENSIONS - AD-4212C-300 / 600 / 3000 / 6000 / 301 / 3100 Unit: mm 9848.83018.846505091.9 1
3111. OPTIONS AX-KO3590-200 RS-232C cable 2 m AX-KO3590-500 RS-232C cable 5 m AX-KO3590-1000 RS-232C cable 10 m (provided with the balanc
32 AD-1684: Electrostatic field meter This option measures the amount of the static charge on the sample, tare or peripheral equipment and display
3312. TERMS/INDEX Terms Stable value The weight value when the stabilization indicator appears. Environment Ambient conditions such as vibration, d
34AX- KO3590-500 ...4, 31 AX- KO3590-1000 ...4, 31 AX-USB-9P...
35Remote controller ...31 Repeatability... 26, 28, 29, 33 RE-ZERO button ...
21. INTRODUCTION This manual describes how the AD-4212C series balance works and how to get the most out of it in terms of performance. Read this
3 Example 1 (Connecting to a personal computer or an AD-8922A) Accessory RS-232C cable (10 m)PersonalcomputerU
4 Example 2 (Connecting to a PLC) . * Specify the AD-8922A-01 connection cable when ordering the AD-8922A-01. AX-KO3705-200 2 m
51.2. Compliance Compliance with FCC Rules Please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has
6Windows communicationtools (WinCT-AD4212C)AC adapterID labelRS-232C cable (10 m)Weighing panPan supportScrews (Flat head: M4 X 6)Calibration weight
7ATTACHING THE AD-4212C SERIES WEIGHING UNIT DIRECTLY ON THE MOUNTING BASE When the AD-4212C series weighing unit is built into a system, remove the