- Initialization
Initialization restores the AD-8922A function settings to factory settings.
1. Connect the AC adapter. The bar display or the weighing data
display appears.
2. While pressing and holding the ON:OFF key, press the PRINT
key. "Clr" appears in the display.
3. Press the PRINT key. (To cancel the operation, press the CAL
key.)
4. Press the RE-ZERO key to select "
go".
5. Press the PRINT key to perform initialization.
After initialization, the bar display or the weighing data display
appears.