3.1. Calibration Group
English
GB 15
The GR series has the following calibration modes and tests.
Automatic Self Calibration
Calibration using the internal weight
Calibraton using an external weight
Calibration test using the internal weight
Calibration test using an external weight
Correction of the internal weight value
Note: Calibration is controlled by the “Permission or Prohibition”.
Refer to Section “Function Switch and Initialization”.
The weight which can be used for calibration test is called “the target weight”. The weight which you
have is called “the external weight”.
Note: It is necessary to perform calibration:
When the balance is installed for the first time.
When the balance has been moved.
When the ambient environment has changed.
For periodical calibration.
Prevent vibration, draughts, and ambient temperature changes during calibration of the balance.
• This indicator means “the balance is measuring calibration data”. Do not allow
vibraton or draughts to affect the balance while this indicator is displayed.
The data for GLP (Good Laboratory Practice) can be output using the RS-232C interface, when the
“GLP output (inFo)” of “Data output (dout)” is set to “/” or “2”. Refer Section 8 “Function table”.
Caution using an External Weight
The accuracy of an external weight can ifluence the accuracy of the balance.
Product Usable external weight Adjustable range
GR-202
200 g, 100g
GR-200 +15.9 mg ~ –15.0 mg
GR-120 100 g, 50 g
GR-300 200 g, 300 g