7. Function Switch and Initialisation
The balance stores parameters that must not be changed carelessly (eg. Calibration data for precision
weighing, Data for adapting to environment, Contro data for RS-232C interface, etc.). There are five
switches for the purpose of preserving these parameters. Each switch can select either “permission” or
“prohibition”. The “prohibition” protects against careless operation.
Switches
(Factory setting)
Function table
0 : To inhibit change to the function table.
1 : To permit change to the function table.
Calibration with the Internal Weight
0 : To inhibit calibration.
1 : To use calibration.
Calibration with an External Weight
0 : To inhibit calibration.
1 : To use calibration.
Automatic Self Calibration
0 : Not used
1 : Used
Correcting the Internal Weight
0 : To inhibit correction.
1 : To use correction.
Operation
1. Turn off the display using the [ON:OFF] key.
2. Press the [ON:OFF] key while the [RANGE] key and [PRINT] key are
pressed and held. Then the balance displays PS.
3. Press the [PRINT] key. Then the balance displays the switch settings.
4. Set the switches using the following keys.
[RE-ZERO] key The key to select the setting of the switch.
[RANGE] key The key to select the switch to change the setting.
[PRINT] key The key to store the new setting.
[CAL] key The key to cancel this operation.
7.1. Permission or Prohibition