A&D GR Series Instruction Manual
GB 32
This function returns the following parameters to factory settings.
Calibration data.
Function table.
The 100% weight.
The data that is stored in the balance using data memory function.
External calibration weight and target weight.
Switch settings for “Permission or Prohibition”.
Operation
1. Turn off the display.
2. Press the [ON:OFF] key while the [RANGE] key and
[PRINT] key are pressed and held. Then the balance displays PS.
3. Press the [RANGE] key to display CLr.
4. Press the [PRINT] key. (If you want to cancel this operation,
press the [CAL] key.)
5. Press the [RE-ZERO] key.
6. Press the [PRINT] key to initialize the balance.
The balance will automatically return to weighing mode.
+
Press and hold
these keys
Press the key
7.2. Initialising the Balance